Applied spectroscopy
Standard reference material 2036 near-infrared reflection wavelength standard
Abstract
Standard Reference Material 2036 (SRM 2036) is a certified transfer standard intended for the verification and calibration of the wavelength/wavenumber scale of near-infrared (NIR) spectrometers operating in diffuse or trans-reflectance mode. SRM 2036 Near-Infrared Wavelength/Wavenumber Reflection Standard is a combination of a rare earth oxide glass of a composition similar to that of SRM 2035 Near-Infrared Transmission Wavelength/Wavenumber Standard and SRM 2065 Ultraviolet-Visible-Near-Infrared Transmission Wavelength/Wavenumber Standard, but is in physical contact with a piece of sintered poly(tetrafluoroethylene) (PTFE). The combination of glass contacted with a nearly ideal diffusely reflecting backing provides reflection-absorption bands that range from 15% R to 40% R. SRM 2036 is certified for the 10% band fraction air wavelength centroid location, (10%)B, of seven bands spanning the spectral region from 975 nm to 1946 nm. It is also certified for the vacuum wavenumber (10%)B of the same seven bands in the spectral region from 10 300 cm(-1) to 5130 cm(-1) at 8 cm(-1) resolution. Informational values are provided for the locations of thirteen additional bands from 334 nm to 804 nm.