Guidelines for transcutaneous p O2 and p CO2 measurement
Abstract
This document provides guidelines in the terminology, methodology, and in the interpretation of data obtained from the use of skin (transcutaneous) p O2 and p CO2 electrodes. The transcutaneous technique has found special application for newborn infants. The causes of analytical bias with respect to arterial blood gas values and imprecision obtained with transcutaneous p O2 and p CO2 electrodes are reviewed. Electrode temperatures above 44 degrees C should not be used routinely, and, at a measuring temperature of 44 degrees C, the measuring site should be changed at least every 4 h to avoid skin burning.
Source and rights
Published by Journal of the International Federation of Clinical Chemistry, 1990. The guideline itself is at the source linked here.
Prognia summarises this guideline for reference and education. It is not medical advice. Always check the official guideline and current local protocols before acting on it.