Impedance based prediction of pulsed field ablation lesion depth without contact force
By Bangjiaxin Ren, Wenjie Li, Fanghui Li and 10 others.
Abstract
The lesion depth prediction index has been demonstrated to optimize ablation strategies. However, contact force-based prediction of lesion depth is difficult to integrate into multi-electrode catheters with complex shapes. This study developed and validated a contact force-independent pulsed-field ablation (PFA) index, the depth prediction index (DPI). Computer modeling identified parameters potentially affecting lesion formation. For in vivo experiments, six swine were assessed 2 h post-ablation to evaluate the difference between DPI and actual lesion depth, and two swine were assessed one month post-ablation to evaluate the difference. DPI predicted 131 of 156 lesions (84.0%) within 1.0 mm of measured depth, with a Pearson correlation coefficient of 0.74, a root mean square error (RMSE) of 0.68 mm, and a mean absolute error (MAE) of 0.56 mm. The DPI–depth association remained significant after adjustment for application number and within-animal clustering (adjusted coefficient, 0.628 mm; 95% CI, 0.505–0.751; P < 0.001). The DPI effectively predicts lesion depth without requiring contact force. This provides proof of feasibility for impedance-based prediction of PFA lesion depth and opens up a new direction for future lesion depth prediction in multi electrode PFA catheters.